Appendix C: Data Port Protocol
9 101112131415
12345678
Figure 14: Data Port Pin Layout
Pins 2, 3, and 5 provide data in RS-232 format.
Pins 1, 4, 9, and 12 provide data in RS-422 format
. TxD+ and
TxD- are the differential transmit data pair
. RxD+ and RxD- are
the differential receive pair.
No hardware flow control is used
. However, in the ASCII mode
XON/XOFF flow control is supported.
BAUD RATE
The baud rate can be changed by pressing the SETUP softkey, the
NEXT softkey, and then the COMM softkey
. Use the Adjust
Up/Down Buttons to select a baud rate of 2400, 9600, or 19200,
depending on the capabilities of the attached equipment.
REAL-TIME DISPLAY FORMAT
While on AC power and while the NPB-295 data port protocol
setting is set to ASCII (refer to Start-Up and Use section),
real-time data is continuously sent to the data port on the back of
the NPB-295
. Patient data can be obtained through the data port
by connecting it to an attached PC or serial printer
. When a
real-time printout or display is being transmitted to a printer or
PC, a new line of data is displayed every 2 seconds
. Column
headings will be displayed or printed after every 25 lines, or if
one of the values in the column heading changes.
Data cannot be obtained if the NPB-295 is operating on battery
power.
Note: If the data output stops transmitting, turn the power off
and back on again or, if connected to a PC, send an XON
(Ctrl-q) to reset the monitor.
An example of a real-time printout is shown in Figure 15.
73