Start-Up and Use
SPORADIC OR LOST PULSE WITH CONTINUOUS MOTION
SPORADIC OR LOST PULSE WITHOUT CONTINUOUS MOTION
Pulse Search Indicator lights continuously
%SpO
2
and Pulse Rate Displays
alternate between previous
display and dashes
Motion Indicator lights continuously
*
0 seconds
10 seconds
Pulse Search
Indicator flashes
Zeros displayed
Low priority
alarm sounds
60 seconds
Pulse Search
Indicator lights
continuously
0 seconds 10 seconds 16 seconds 60 seconds
%SpO
2
and Pulse
Rate Displays alternate
between previous
display and dashes
%SpO
2
and pulse rate flashes zeros
Pulse Search Indicator flashes
High priority alarm sounds
If during motion the NPB-295 detects a qualified pulse, the values on the display are updated.
If motion stops at any time during the pulse search mode, and no qualified pulse is detected,
the high priority alarm will immediately sound after 6 seconds.
*
50 seconds
6 seconds
50 seconds
Figure 7: Monitoring Conditions and Alarm Responses
The NPB-290 manages loss-of-pulse alarms as follows, and as
illustrated in Figure 7. If the NPB-290 fails to detect at least one
qualified pulse during any 10-second period, it enters pulse
search mode, the Pulse Search indicator lights, the displays
alternate between data and dashes, and the data evaluation period
starts.