TROUBLESHOOTING AND MAINTENANCE
Troubleshooting
EMI Interference
Obtaining Technical Assistance
Returning the NPB-290
Maintenance
TROUBLESHOOTING
WARNING: if you are uncertain about the accuracy of any
measurement, check the patient’s vital signs by alternate
means; then make sure the monitor is functioning correctly.
WARNING: The cover should be removed only by qualified
service personnel. There are no user-serviceable parts inside.
If you experience a problem while using the NPB-290 and are
unable to correct it, contact qualified service personnel or
Mallinckrodt’s representative. The NPB-290 service manual,
which is for use by qualified service personnel, provides
additional troubleshooting information.
Following is a list of possible errors and suggestions for
correcting them.
1. There is no response to the Power On/Off button.
• If operating on AC power, ensure that the supply voltage
selector switch is set to the proper voltage.
• If operating on AC power, the fuse may be blown.
Notify service personnel to check and, if necessary,
replace the fuse.
• If operating on battery power, the battery may be missing
or discharged. If the battery is discharged, notify service
personnel to charge or replace the battery.
2. One or more display segments or indicators do not light
during the power-on self-test.