Mallinckrodt NPB-190 Instrukcja Operatora Strona 101

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 110
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 100
Data Port Protocol
93
Patient Data and Operating Status
Time
TIME %SpO2 BPM PA Status
12-NOV-99 14:00:05 100 120 50
The Time column represents the N-395 real-time clock.
Patient Data
N-395 VERSION 2.0.0.0 CRC: XXXX SpO2 Limit: 70-100% PR Limit: 60-160BPM
TIME %SpO2 BPM PA Status
12-NOV-99 14:00:05 100 120 50
Patient data is highlighted in the display above. Parameter values
are displayed directly beneath the heading for each parameter. In
this example the %SpO
2 is 100, and the pulse rate is 190 beats
per minute. The “*” next to the 190 indicates that 190 beats per
minute is outside of the alarm limits, indicated in the top row, for
pulse rate. If no data for a parameter is available three dashes
(- - -) will be displayed.
PA is an indication of pulse amplitude. The number can range
from 0 to 254. There are no alarm parameters for this value. It
can be used for trending information as an indication of a change
in pulse volume, relative pulse strength, or circulation.
Operating Status
N-395 VERSION 2.0.0.0 CRC: XXXX SpO2 Limit: 70-100% PR Limit: 60-160BPM
TIME %SpO2 BPM PA Status
12-NOV-99 14:00:11 100 190* 50 PS SL PH LB
The Status column indicates alarm conditions and operating
status of the N-395. In this example, the PH means that the pulse
rate upper alarm limit (Pulse High) has been exceeded. A
complete listing of the status codes is listed in Table 11. As
many as 4 codes can be displayed at one time in the Status
column.
Przeglądanie stron 100
1 2 ... 96 97 98 99 100 101 102 103 104 105 106 ... 109 110

Komentarze do niniejszej Instrukcji

Brak uwag