Introduction
4
Bone, tissue, pigmentation, and venous vessels normally absorb a
constant amount of light over time. The arteriolar bed normally
pulsates and absorbs variable amounts of light during the
pulsations. The ratio of light absorbed is translated into a
measurement of functional oxygen saturation (SpO
2).
Note: For an explanation of functional versus fractional
saturation, refer to Principles of Operation.
Because a measurement of SpO
2 is dependent upon light from the
sensor, excessive ambient light can interfere with this
measurement.
Specific information about ambient environmental conditions,
sensor application, and patient conditions is contained throughout
this manual.