Data Port Protocol
90
Note: Selections for Agilent, SpaceLabs, GE Marquette, and
Datex are included in N-395 software versions 1.7 and
higher.
The protocol settings allow the N-395 to communicate with
various devices. ASCII is used during normal operation and for
serial printers. GRAPH is used for graphical trend printouts
when connected to a serial printer that supports Epson ESC
protocol. Select OXINET when connecting to an Oxinet
â
II
central station network or Score software. The Bedside Monitor
Interface selections are AGILENT (for Agilent HP monitors),
SPACELBS (for SpaceLab monitors), MARQ (for Marquette
monitors), and DATEX (for Datex-Ohmeda AS/3 monitors). Do
not use the CLINICAL setting unless instructed to do so by
Mallinckrodt’s Technical Services personnel.
REAL-TIME DISPLAY FORMAT
While the N-395 data port protocol setting is set to ASCII (refer
to Start-Up and Use section), real-time data is continuously sent
to the data port on the back of the N-395. Patient data can be
obtained through the data port by connecting it to an attached PC
or serial printer. When a real-time printout or display is being
transmitted to a printer or PC, a new line of data is displayed
every 4 seconds. Column headings will be displayed or printed
after every 25 lines, or if one of the values in the column heading
changes.
Note: If the data output stops transmitting, turn the power off
and back on again or, if connected to a PC, send an XON
(Ctrl-q) to reset the monitor.
An example of a real-time printout is shown in Figure 23.