Data Port Protocol
91
N-395 VERSION 2.0.0.0 CRC: XXXX SpO2 Limit: 70-100% PR Limit: 60-160BPM
TIME %SpO2 BPM PA Status
12-NOV-99 14:00:05 100 120 50
12-NOV-99 14:00:07 100 124 50
12-NOV-99 14:00:09 100 190* 52 PS SL PH LB
12-NOV-99 14:00:11 100 190* 50 PS SL PH LB
12-NOV-99 14:00:13 100 190* 51 PS SL PH LB
12-NOV-99 14:00:15 100 190* 50 PS SL PH LB
12-NOV-99 14:00:17 100 190* 50 PS SL PH LB
12-NOV-99 14:00:19 100 190* 51 PS SL PH LB
12-NOV-99 14:00:21 100 190* 53 PS SL PH LB
12-NOV-9914:00:23 100 190* 50 PS SL PH LB
12-NOV-99 14:00:25 100 190* 50 PS SL PH LB
12-NOV-99 14:00:27 - - - - - - - - - SD SL PH LB
12-NOV-99 14:00:29 - - - - - - - - - SD SL PH LB
12-NOV-99 14:00:31 - - - - - - - - - SD SL PH LB
12-NOV-99 14:00:33 - - - - - - - - - SD
12-NOV-99 14:00:35 - - - - - - - - - SD
12-NOV-99 14:00:37 - - - - - - - - - SD
12-NOV-99 14:00:39 - - - - - - - - - SD
12-NOV-99 14:00:41 - - - - - - - - - SD
12-NOV-99 14:00:43 - - - - - - - - - SD
12-NOV-99 14:00:45 - - - - - - - - - SD
12-NOV-99 14:00:47 - - - - - - - - - SD
12-NOV-99 14:00:49 - - - - - - - - - SD
N-395 VERSION 2.0.0.0 CRC: XXXX SpO2 Limit: 70-100% PR Limit: 60-160BPM
TIME %SpO2 BPM PA Status
12-NOV-99 14:00:51 - - - - - - - - - SD
N-395 VERSION 2.0.0.0 CRC: XXXX SpO2 Limit: 70-100% PR Limit: 60-160BPM
TIME %SpO2 BPM PA Status
12-NOV-99 14:00:53 79* 59* 50 PS SL PL LB
12-NOV-99 14:00:55 79* 59* 50 PS SL PL LB
Figure 23: Real-Time Printout
Column Headings
Every 25
th
line of the data is a column heading.
N-395 VERSION 2.0.0.0 CRC: XXXX SpO2 Limit: 70-100% PR Limit: 60-160BPM
TIME %SpO2 BPM PA Status
A column heading is also displayed whenever a value of the
column heading is changed. There are three column heading
lines shown in Figure 23. Using the top row as the starting point
there are 25 lines before the second column heading is printed.
The third column heading was displayed because the SpO
2 limits
changed from 70-100% to 80-100%.