Specifications
74
Pulse Rate
1
Without Motion 20 - 250 ±3 digits
With Motion Normal physiologic range (e.g., 55 – 125
bpm) ±5 digits
ELECTRICAL
Instrument
Power Requirements 100 - 120 VAC, 200 - 240 VAC,
50/60 Hz, 20 VA switch selectable
Fuses 2 qty, 0.5 A, 250 volts, slow-blow,
IEC (5 x 20 mm)
Battery
Type Lead-Acid
Battery Capacity
A minimum of 2 hours with a new, fully charged battery under
the following conditions: no alarms, and no analog or serial
output devices attached. A completely discharged battery can be
fully recharged in approximately 14 hours while unit is turned
off or 18 hours while turned on.
Charge/discharge cycles: at least 400
Sensor
The wavelength range of the light emitted is within the range of
250 nm to 1,000 nm with the energy not exceeding 10 mw.
1
Pulse Rate accuracy is expressed as plus or minus “X” digits (bpm) across the display range.
This variation equals ± one standard deviation (1SD), which encompasses 68% of the population.