
Start-Up and Use
35
SatSeconds Alarm Management
With traditional alarm management, upper and lower alarm
limits are set for monitoring oxygen saturation. During
monitoring, as soon as an alarm limit is violated by as little as
one percentage point, an audible alarm immediately sounds.
When the %SpO
2 level fluctuates near an alarm limit, the alarm
sounds each time the limit is violated. Such frequent alarms can
be distracting.
The N-395 utilizes Nellcor SatSeconds alarm management
technique. With the SatSeconds technique, upper and lower
alarm limits are set in the same way as traditional alarm
management. The clinician also sets a SatSeconds limit that
allows monitoring of %SpO
2 below the selected low alarm limit
for a period of time before an audible alarm sounds.
The SatSeconds limit controls the time that the %SpO
2 level
may fall below the alarm limit before an audible alarm sounds.
The method of calculation is as follows:
The number of percentage points that the %SpO
2 falls outside
the alarm limit is multiplied by the number of seconds that the
%SpO
2 level remains outside that limit. This can be stated as an
equation:
Points x Seconds = SatSeconds
Where:
Points = SpO
2 percentage points below of the limit
Seconds = number of seconds SpO
2 remains at
that point below of the limit
For example, Figure 11 demonstrates the alarm response time
assuming a SatSeconds limit set at 50, and a lower alarm limit
set at 90 percent.
In this example, the %SpO
2 level drops to 88 (2 points) and
remains there for a period of 2 seconds (2 points x 2 seconds =
4). The %SpO
2 then drops to 86 for 3 seconds and then to 84
for 6 seconds. The resulting SatSeconds are:
Komentarze do niniejszej Instrukcji