Start-Up and Use
55
Figure 20: SpaceLabs Connection
Caution: The SpaceLabs monitor must be turned on before
the N-395 monitor is turned on.
The SpaceLabs monitor requires a Universal FlexPort
interface module to interface with the N-395 pulse oximeter.
The RS-232 hardwire interface cable has a DB-15 connector for
the N-395 and the applicable connector for the SpaceLabs
FlexPort interface module cable. Nellcor cable part number
036341 is recommended for this interface.
Note: Spare parts for the N-395 are listed on the Internet,
confirm the cable part number before ordering a cable.
The Internet address is:
http://mallinckrodt.com/respiratory/resp/Serv_Supp/Apa
rtweb/main/PartAcceMenu.html.
Corrupt data will be indicated by a Communications Error
displayed on the SpaceLabs monitor. The SpaceLabs monitor
will detect corrupt data in less than 11 seconds.
When the N-395 is in the SpaceLabs mode of operation the
interface baud rate must be set to 9,600 bits per second.
Pressing the SETUP softkey, then the NEXT softkey, and then
the COMM softkey to select BAUD. Use the ADJUST UP or
ADJUST DOWN buttons to select the correct baud rate.