Start-Up and Use
42
A description of each menu item is included in the following
paragraphs.
LIMITS
WARNING: Each time the monitor is used, check alarm
limits to ensure that they are appropriate for the patient
being monitored.
Overview
When the N-395 is first turned on, alarm limits are set to their
power-on default values. Qualified service personnel, using the
instructions described in the N-395 service manual, may change
power-on default alarm limits.
Alarm limits may be changed from their power-on default values
if necessary, as described below. Limit changes made will
remain in effect until changed again or until the N-395 is turned
off.
Viewing Current Alarm Limits
To view the current alarm limit values from the main menu,
press the LIMITS softkey. The current upper and lower alarm
limits for %SpO
2 and pulse rate are displayed. The current
SatSeconds alarm limit is also shown.
Przeglądanie stron 49