Start-Up and Use
28
WARNING: Each time the monitor is used, check alarm
limits to ensure that they are appropriate for the patient
being monitored.
Important! Prior to using the N-395, carefully read this manual,
accessory directions for use, all precautionary information in
boldface type, and all specifications.
Before using the N-395 in a clinical setting, verify that the
monitor is working properly and is safe to use. Proper working
condition can be verified by successful completion of the power-
on self-test described in the following steps, and by following
instructions contained in the “Monitoring Mode” paragraph of
this section.
Ensure that the supply voltage selector switch on the rear panel
matches the AC voltage at your location.
Power-On Self-Test (POST)
WARNING: Ensure that the speaker is clear of any
obstruction. Failure to do so could result in an inaudible
alarm tone.
1. Plug an appropriate Nellcor sensor firmly into the SCP-10
or MC-10 cable and lower the SCP-10 or MC-10 sensor
lock over the sensor connector until it clicks into place.
Insert the SCP-10 or MC-10 into the N-395 sensor port.
Apply the sensor to the patient as described in the sensor
directions for use.
2. Turn on the N-395 by pressing the POWER ON/OFF
Button. The monitor automatically starts a power-on self-
test (POST), which tests its circuitry.